Hi Jed,
I'm wondering if you have heated the KOH. For an etch rate of about 80 µm/h
you'll need to heat the etchant to about 70-80 °C.
Regards
Fabio Tagliareni
-----Ursprüngliche Nachricht-----
Von: [email protected]
[mailto:[email protected]]Im
Auftrag von Jed Ley
Gesendet: Mittwoch, 18. Februar 2004 04:00
An: [email protected]
Betreff: [mems-talk] <100> Si Etch rates in KOH?
Hi.
I am trying to etch 100um deep V-pits in on 1-0-0 Si wafers.
I am etching n-type (phosphorous doped) wafers of resistivity
7.5 - 12.5 ohm-cm. I'm using a 10-weight % aqueous solution of
KOH (i.e. 50g KOH pellets + 450 mL D.I. water) , and a ~1um-thick
thermal oxide as an etch mask. The etch pits are defined by 120
um square holes in the oxide. I am getting an etch rate of only
~10 um/ hr for <100> Si. From the literature I have read (Seidel,
et al. J. Electrochem. Soc. V. 137, No.11, Nov 1990), I expected an etch
rate of ~80 um/hr. I am getting an etch rate for the oxide
that is close to what I expected (800-900 Angstrom/hr). Can
anybody suggest to me what I'm doing wrong? Any suggestion will
be greatly appreciated.
Thanks,
Jed Ley
Colorado School of Mines
Golden, Colorado